57-649 Jupiter Magnetostrictive Transmitter - FOUNDATION fieldbus™
27
5.3 Specifications
5.3.1 Functional
System Design
Measurement Principle Magnetostrictive time-of-flight
Input
Measured Variable A return signal is generated from the precise location where the magnetic
field of the MLI float intersects the magnetostrictive wire
Z
ero and Span 6 inches to 400 inches (15 to 999 cm)
User Interface
Keypad 3-button menu-driven data entry and system security
Indication 2-line × 8-character display
Digital Communication F
OUNDATION fieldbus
, H1 (31.25 kbits/sec)
Interoperability test kit (ITK Revision) ITK 4.61
LAS capable Yes, Device type: Linkmaster
Minimum Operating Voltage 9 VDC
Quiescent Current Draw 15 mA
DEV Revision 0X01
Function Blocks 1xRB, 2xAI(s)
Damping Adjustable 0-25 seconds
Power (Measured at instrument terminals)
Fieldbus General Purpose/Explosion Proof 9 to 32 VDC (17 mA maximum current draw)
FISCO/FNICO 9–17.5 VDC (17 mA maximum current draw)
Housing
Material Aluminum A356T6 (<0.20% copper), optional 316 stainless steel
Cable Entry
3
4" NPT and M20
5.3.2 Performance
Accuracy ±0.015"
Repeatability ±0.005% of full span or 0.005" (0.127 mm) (whichever is greater)
Linearity 0.020% of full span or 0.031" (0.794 mm) (whichever is greater)
Maximum level rate of change 6 inches per second
Response time 0.1 second
Warm-up <5 second
Upper dead zone None
Lower dead zone <2" (5 cm), SIL 2: <5" (13 cm)
Ambient temperature range Transmitter: -40° to +175° F (-40° to +80° C)
LCD: -10° to +160° F (-20° to +70° C)
Process temperature External Mount: -40° to +248° F (-40° to +120° C)
-320° to +850° F (-195° to +455° C) (with factory insulated MLI)
Direct Insertion: -40° to +200° F (-40° to +95° C)
High temperature probe: -40° to +500° F (-40° to +260° C)
Humidity 0 to 99% non-condensing
Electromagnetic compliance EN 61326
Maximum Pressure (Direct Insertion) 1700 psig @ +100° F